ITC Hemochron Signature Elite Whole Blood Microcoagulation System User Manual

Page 22

Advertising
background image

22

Specifying Continuous POCT-1A Communication

Note: Communication is initiated from the instrument only when the CLSI POCT-1A compliant
communication protocol is used. If the ITC proprietary communication protocol is used,
communication is initiated from a personal computer or network.

1.

From the Supervisor menu, press 7 to display the Communications menu:

2.

Press 3 to select the desired communication option when the CLSI POCT-1A compliant
communication protocol is used. (Display OFF to initiate communication only if requested, or
display ON to initiate continuous communication whenever the instrument is started.)

Note: Specifying the communication option acts as a toggle. If initiation of communication upon
startup is already specified (Push POCT ON is displayed), this can be cancelled by pressing 3 again
to display Push POCT OFF.

Entering Network Communication Parameters

Some or all of the following parameters may need to be entered in order for the instrument to
communicate over a network:

IP Address

Gateway

NetMask

Remote IP

Remote Port

Note: Consult with the network administrator for your facility to determine which parameters
must be entered and what those settings should be.

1.

From the Supervisor menu, press 7 to display the Communications menu:

2.

(If the CLSI POCT-1A compliant communication protocol is currently selected): Press 1 to toggle the
communication protocol to the ITC proprietary communication protocol:

Note: The CLSI POCT-1A compliant communication protocol must be OFF when running XPORT
Setup.

3.

Press 4 to start the XPORT Setup module. The message XPORT Setup Please Wait is displayed,
followed by the first page of the XPORT Setup module:

Note: If the message XPORT Setup NOT RESPONDING! is displayed, contact ITC Technical Support
at (800) 631-5945 or (732) 548-5700, or by e-mail at [email protected].

Advertising