Error Message

Cause

Corrective Action

Detector Fault

Light path between LED and
detectors is blocked.

Repeat test with new cuvette. If
message persists, contact ITC
Technical Support.

Detector Blocked

EQC is being run while a cuvette
is inserted.

Remove the cuvette and repeat
the EQC.

Sample Pos Fault

Sample has moved outside of
testing area in cuvette.

Contact ITC Technical
Support.

ASSAY LOCKED
CALL SUPERVISOR

The QC failure limit has been
exceeded.

Remove the cuvette. Contact
the supervisor.

MACHINE LOCKED
CALL SUPERVISOR

The EQC failure limit has been
exceeded.

Contact the Supervisor.

Sample Not Seen

Sample has not reached front
detector in specified time period.

Repeat test with new cuvette. If
message persists, contact ITC
Technical Support.

Cuvette Removed

Cuvette was prematurely removed
from instrument while testing in
progress.

Repeat test with new cuvette.

Sample Too Large/
Sample Too Small

Excess or insufficient sample.

Repeat test with new cuvette.

Premature Sample

Sample was detected at front
detector before specified time
period. May occur if sample is
added before pump-priming
sequence is complete.

Repeat test with new cuvette. If
message persists, contact ITC
Technical Support.

MEMORY FAULT

Malfunction in the computer's
memory.

Contact ITC Technical
Support.

START Timed Out

START key was not pressed within
5 minutes after entering ready
mode.

Remove cuvette and repeat test
with new cuvette.

Unsupported Assay

Cuvette cannot be identified by
instrument.

Remove cuvette and repeat test
with new cuvette. Use ITC
cuvettes only.

No Data Stored

Operator attempted to print data
when none is stored in database.

N/A

User Abort

The test was aborted by the user.

Repeat test.

Invalid Lot #

The barcoded cuvette or QC lot
number label that was scanned
was not recognized.

Check the label for damage,
then repeat the scan.

The incorrect format was entered
for the lot number.

Repeat the entry using the
correct format.

The lot that was scanned does not
match the cuvette test type.

Verify the test type, then rescan
the lot.