Safety features – CDI Torque MULTITEST Motorized Loader User Manual

Page 12

Advertising
background image

IV

Safety Information

safety features

LIMITING SWITCHES

The limiting switches on the loading mechanism will automatically stop the motor
before damage is done to the loading mechanism. To restart the motor reverse the
direction of the load.

“ OVER TORQUE ” PROTECTION FOR TRANSDUCER IN USE

The motor will stop when the transducer in use has reached an “OVER RANGE”
condition. This protects the transducer from permanent damage.

INPUT WHEEL ENGAGE/DISENGAGE POSITION

The fine adjustment input wheel is to allow the operator to apply load in small
increments when testing DIAL indicating, DEFLECTING beam and ELECTRONIC
TORQUE wrench. When not in use it has a disengage feature to prevent injury to
the operator.

IT IS RECOMMENDED TO DISENGAGE INPUT WHEEL PRIOR TO ACTIVATING
THE MOTOR.

To disengage the input wheel, pull the wheel out until it is disengaged from the
loading mechanism. When disengaged, with the motor activated, the input wheel
will rotate slightly but will ridding free on the input shaft.

To engage the input wheel for use, press the “EMERGENCY STOP” button on the
control box to disengage the motor. Push the wheel in until it engages with the
input shaft.

Advertising