Downloading data list to personal computer – CDI Torque MULTITEST Motorized Loader User Manual

Page 56

Advertising
background image

4-12

Using the Multitest Toque Calbration System

5. Press ENTER to download the raw data list. The LED

display reads:

SENDING

6. Press ENTER again to return to measurements mode.

Downloading Data List to Personal Computer

1. Connect the accessory 2000-50-1 cable between the

2000-810-01 PRINTER port and the computer COM port.
For additional information on setting up printer or PC
ports, refer to Chapter 3—Set Up and Programming .

2. Set up your computer to receive serial communication.

Example:

If using an IBM compatible and Windows 95:

Click on START—RUN to open the Terminal
program.

Type TERMINAL.

Press <Enter>.

3. Press RECALL DATA.

4. Press SEND/AUTO SEND. As shown in Figure 4-5, the

LCD display reads:

Download - upper line

+Date Time Raw - lower line, with the plus sign flashing

To download the data list plus date-time stamp,
press ENTER. The LED display reads:

SENDING

To download the raw data list, use right DIGIT to
select “Raw” on the LCD display. Then press
ENTER to download. The LED display reads:

SENDING

5. Press ENTER again to return to measurements mode.

6. Copy the data list to your chosen software.

Example:

If using an IBM compatible and Windows 95:

Select EDIT —SELECT ALL, and then EDIT—COPY.

Open the spreadsheet or data program being used.

Select EDIT—PASTE in the program being used.

Advertising