Angle only – CDI Torque MULTITEST Motorized Loader User Manual

Page 59

Advertising
background image

4-15

Using the Multitest Toque Calbration System

Angle Only

The 2000-810-01 accepts input from common bi-directional
incremental rotary encoders. The Indicator interfaces to any
encoder resolution (counts per revolution) and can be
programmed to display angle in degrees of rotation.

For additional information on setting up angle encoders and
measurement modes, refer to Chapter 3—Set Up and
Programming
in this manual.

1. Press MODE to select ANGLE.

LED displays torque.

LCD displays degrees of rotation at the lower left.

Figure 4-6: Angle Mode Display

2. Install the angle encoder between the fastener and an

appropriate wrench or breaker bar. Fix the encoder
reference arm, strap, clamp or magnet to the work.

3. Press ANGLE RESET to zero the angle display.

4. Rotate the fastener while monitoring the angle on the

LCD display. The SET HIGH limit momentarily sounds
an audible tone at angle preset coincidence (when the
measured angle matches the SET LOW angle).

5. Press ANGLE RESET to clear the angle display.