Quick check, Warning – CDI Torque MULTITEST Motorized Loader User Manual

Page 65

Advertising
background image

5-5

Calibration

Quick Check

WARNING

Always be alert to the potential for personal injury
that may be caused by excessive torque
applications, careless handling of heavy weights,
and out-of-balance or unsafe weight distribution.

1. With the 2000-810-01 Indicator ON, select TORQUE

measurement, TRACK mode, and select the UNITs of
measure appropriate to the transducer as specified in the
calibration check point table in this chapter.

2. Install the transducer to be checked and allow a 5 minute

warm up period to stabilize the electronic circuits and
transducer elements.

3. Transducers are mounted horizontally for calibration as

instructed in Mounting Details in this chapter. Mount the
4-in-1 transducer to a sturdy bench using the bracket
provided. Mount single transducers horizontally using
the accessory stand and block. Attach the specified
torque test bar, wheel or butterfly, secured with its drive
retention screw, and install the specified hanger for
clockwise (CW) torque application.

4. Apply and remove the full range torque for the transducer

three times.

5. Remove the weights and hanger, but leave the arm

attached, and press ZERO TARE.

6. At each of the check points specified for the transducer

in the calibration check point table (in this chapter), apply
torque in the CW direction and verify the display
readings are within 0.25% of applied torque. If any
reading is out of specification, proceed to Calibrating
MULTITEST Torque/Force Transducers
in this chapter.

7. Repeat Steps 3 through 6 in the counter clockwise

(CCW) direction.

When using the 2-1/2" or 5" test wheel, a small “S”
hook p/n 2000-151-4 is supplied to provide a positive
zero bias in the direction (CW or CCW) that the
calibration is performed. Do not remove the “S” hook
when setting zero tare.

Advertising