10.8 Data Formats

Continuous Output Serial Data Format

If continuous transmission is configured for a serial
port (STREAM parameter set to LFT or INDUST on
the SERIAL menu), the

920i

sends data using the

Consolidated Controls serial data format shown in
Figure 10-2:

Figure 10-2. Continuous Output Serial Data Format

Demand Output Serial Data Format

When demand mode is configured for the serial port
(STREAM parameter set to OFF), the

920i

uses a data

string formatted for a basic ticket printout. The
particular ticket format printed depends on the
indicator configuration.

You can customize the ticket to work with a wide
variety of printers, scoreboard displays, and other
remote equipment. See Section 6.0 on page 57 for
more information on custom print formats.

RS-485 Data Formats

Two-wire RS-485 communications is available on
p o r t 3 o f t h e C P U b o a r d ; f o u r- w i r e R S - 4 8 5
communications is supported on the “A” ports of any
installed serial expansion cards.

The

920i

has a built-in RS-485 software protocol

which is enabled when you assign a non-zero address
to the indicator. Valid RS-485 addresses must be in
the range 1–255; the address is specified on the
ADDRESS parameter on the SERIAL menu.

All remote commands are initiated using the data
format shown in Figure 10-3:

Figure 10-3. RS-485 Send Data Format

If the initiating device address matches the port address
of an

920i

on the RS-485 network, that indicator

responds. For example, with demand outputs, or in
response to a KPRINT command, the responding
indicator uses the format shown in Figure 10-4:

Figure 10-4. RS-485 Respond Data Format

Example:

To send the KPRINT command from an

ASCII terminal to an indicator at address 65 (decimal)
on the RS-485 network, use the format shown in
Figure 10-3.

The keyboard equivalent for the start-of-text
(

STX

) character is

CONTROL-B

.

The indicator address (65) is represented by an
upper case “A”.

The carriage return (CR) character is generated
by pressing the

ENTER

key.

Therefore, to send the KPRINT command to the
indicator at address 65, enter the following at the
terminal:

CONTROL-B

, A, K, P, R, I, N, T,

ENTER

.

The indicator responds with the format shown in
Figure 10-4:

<STX> A

SCALE #1

<EOL>

GROSS 1699 LB<EOL>
08/20/1998 10:05 AM<EOL>

<ETX> <CR>

If continuous transmission is configured for the serial
port, the

920i

sends data using the data format shown

in Figure 10-5:

Figure 10-5. RS-485 Continuous Data Format

<STX> <POL> <wwwwwww> <UNIT> <G/N> <S> <TERM>

ASCII 02

decimal

Polarity:
<space> = Positive
<–> = Negative

Weight: 7 digits, right-justified, dummy
zeroes, decimal point with no leading
zeroes except for leading zero immediately
preceding the decimal point. Leading
zeroes transmitted as spaces.

L = pounds
K = kilograms
T = tons
G = grains
<space> = grams
O = ounces

G = Gross

N = Net

Status:
<space> = valid
I = Invalid
M = Motion
O = Over/under range

<CR> <LF>

or <CR>

<STX> <ADDRESS> <COMMAND> <CR>

ASCII 02

decimal

Address of the

receiving indicator

EDP serial command

NOTE: Host must send <CR>, not

<CR><LF>. Failure to use <CR>
renders all indicators unable to
respond to serial commands

ASCII

13 decimal

<STX> <ADDRESS> <DEMAND DATA> <ETX> <CR>

ASCII 02

decimal

Address of the

transmitting indicator

Response commands from
indicator:
<first line > <EOL>
<additional lines> <EOL>
<last line> <EOL>
where EOL = <CR> or <CR> <LF>
(set by TERMIN parameter on the
SERIAL menu)

ASCII 13

decimal

ASCII 03

decimal

Response commands from
indicator:
<STX> <POL> <wwwwwww>
<L/K> <G/N> <S> <EOL>
NOTE:
This is the Consolidated

Controls continuous data format

<STX> <ADDRESS> <STREAM DATA> <ETX> <CR>

ASCII 02

decimal

Address of the

transmitting indicator

ASCII 13

decimal

ASCII 03

decimal