0 configuration, 1 configuration methods, 1 irev configuration – Rice Lake 920i Installation Manual V3.07 User Manual

Page 25: Configuration, Irev configuration

Advertising
background image

Configuration

19

3.0

Configuration

To configure the

920i

indicator, the indicator must be

placed in setup mode. The setup switch is accessed by
removing the large fillister head screw on the
backplate of the universal and deep enclosures.
Switch position is changed by inserting a screwdriver
into the access hole and pressing the switch.

When the indicator is placed in setup mode, a series of
menus is shown across the top of the display, along
with the words

Scale Configuration

. The SCALES menu

is highlighted as the first used to configure the
indicator. Detailed descriptions of these menus are
provided in Section 3.2.

When configuration is complete, press the

Save and

Exit

softkey to exit setup mode, then replace the setup

switch access screw.

3.1

Configuration Methods

The

920i

indicator can be configured by using the

front panel keys to navigate through a series of
configuration menus or by sending commands or
configuration data to an indicator serial port.
Configuration using the menus is described in
Section 3.1.3.

C o n f i g u r a t i o n u s i n g t h e s e r i a l p o r t c a n b e
accomplished using either the serial command set
described in Section 9.0 or the

iRev

configuration

utility.

NOTE:

Some configuration parameters, such as those

used to configure the

920i

display and widgets, cannot

be accessed through the configuration menus.

iRev

p r o v i d e s t h e m o s t c o m p l e t e a n d e f f i c i e n t
configuration interface for the

920i

.

3.1.1

iRev Configuration

The

iRev

configuration utility provides the preferred

method for configuring the

920i

indicator.

iRev

runs

o n a p e r s o n a l c o m p u t e r t o s e t c o n f i g u r a t i o n
parameters for the indicator. When

iRev

configuration

is complete, configuration data is downloaded to the
indicator.

iRev

supports both uploading and downloading of

indicator configuration data. This capability allows
configuration data to be retrieved from one indicator,
edited, then downloaded to another indicator with an
identical hardware configuration.

Figure 3-1. iRev Hardware Configuration Display

To use

iRev

, do the following:

1. Install

iRev

on an IBM-compatible personal

computer. See Section 5.0 on page 54 for
detailed hardware and software requirements.

2. With both indicator and PC powered off,

connect the PC serial port to the RS-232 pins
on the indicator serial port.

3. Power up the PC and the indicator. Use the

setup switch to place the indicator in setup
mode.

4. Start the

iRev

program.

iRev

provides online help for each of its configuration

displays. Parameter descriptions provided in this
manual for front panel configuration can also be used
when configuring the indicator using

iRev

: The

interface is different, but the parameters set are the
same.

See Section 5.0 on page 54 for more information about
using

iRev

to configure the

920i

.

Advertising