Serial Commands

85

9.1.2

Reporting Commands

Reporting commands send specific information to the
serial port. The commands listed in Table 9-2 can be
used in either setup mode and normal mode.

9.1.3

Clear and Reset Commands

The following commands can be used to clear and
reset the

920i

:

PCLR

: Program clear. Erases the loaded user

program (setup mode only).

RS

: Reset system. Resets the indicator without

resetting the configuration.

RESETCONFIGURATION

: Restores all configuration

parameters to their default values (setup mode
only). The

RESETCONFIGURATION

function can

also be initiated by pressing the

Reset Config

softkey under the VERSION menu. Use the

up

key to select

Clear entire indicator configuration

, then

press

ENTER

to reset the indicator.

NOTE:

All load cell calibration settings are lost when the

RESETCONFIGURATION command is run.

9.1.4

Parameter Setting Commands

Parameter setting commands allow you to display or
change the current value for a particular configuration
parameter (Tables 9-3 through 9-12).

Current configuration parameter settings can be
displayed in either setup mode or normal mode using
the following syntax:

command<ENTER>

Most parameter values can be changed in setup mode
only; setpoint parameters listed in Table 9-6 on
page 89
can be changed when in normal weighing
mode.

Use the following command syntax when changing
parameter values:

command=value<ENTER>

, where

value

is either a number or a parameter value. Use no spaces
before or after the equal (=) sign. If you type an
incorrect command, the display reads

??

.

For example, to set the motion band parameter on
Scale #1 to 5 divisions, type the following:

SC.MOTBAND#1=5D<ENTER>

For parameters with selectable values, enter the
command and equal sign followed by a question
mark:

command=?<ENTER>

to see a list of those values.

The indicator must be in setup mode to use this
function.

NOTE: Some parameters are valid only if other parameters or
parameter values are specified. See the configuration menus
in Section 3.2 on page 21 for inform
ation about parameter
dependencies. Restrictions for front-panel configuration also
apply to serial command configuration.

Command

Function

DUMPALL

List all parameter values

SPDUMP

Print setpoint configuration

VERSION

Write

920i

software version

HARDWARE

Lists option cards installed in slots 1–14.
See Section 10.1.2 on page 102 for more
information about using the HARDWARE
command.

XE

Returns a10-digit code representing any
error conditions currently shown on the
front panel. See Section 10.1.4 on
page 103 for mor
e information.

Table 9-2. Reporting Commands

Command

Description

Values

SC.GRADS#n

Graduations

1–9999999

SC.SPLIT#n

Multi-range or multi-interval
scale type

OFF, 2RNG, 3RNG, 2INTVL, 3INTVL

SC.ZTRKBND#n

Zero track band

0, 0–100

SC.ZRANGE#n

Zero range

1.900000, 0–100

SC.MOTBAND#n

Motion band

1, 0–100

SC.SSTIME#n

Standstill time

1–65535

SC.OVRLOAD#n

Overload

FS+2%, FS+1D, FS+9D, FS

SC.WMTTHRH#n

Weighment threshold

grads

SC.NUMWEIGH#n

Number of weighments

SC.MAX_WEIGHT#n

Maximum weight

Table 9-3. SCALES Serial Commands