3 serial command calibration – Rice Lake 920i Installation Manual V3.07 User Manual

Page 57

Advertising
background image

Calibration

51

Figure 4-2. WZERO Calibration Display

To calibrate the indicator using the front panel, do the
following:

1. Place the indicator in setup mode (display

reads

Scale Configuration

) and remove all

weight from the scale platform. If your test
weights require hooks or chains, place the
h o o k s o r c h a i n s o n t h e s c a l e f o r z e r o
calibration.

2. With the

SCALES

menu highlighted, press the

d o w n

k e y, t h e n s e l e c t t h e s c a l e t o b e

c a l i b r a t e d . P r e s s

d o w n

a g a i n (

G R A D S

parameter highlighted), then press

left

to

h i g h l i g h t t h e

C A L I B R

s u b m e n u ( s e e

F i g u r e 4 - 1 ) . P r e s s

d o w n

t o g o t o z e r o

calibration (

WZE RO

). See Figure 4-2 on

page 51.

3. Ensure scale is empty, then press

down

again

to show the current WZERO value, then press
the

Calibrate

softkey to calibrate zero. When

complete, the new A/D count for the zero
calibration is displayed. Press

enter

to save the

zero calibration value and go to the next
prompt (

WVAL

).

4. With

WVAL

displayed, press

down

to show the

stored calibration weight value. Use the
numeric keypad to enter the actual value of
the calibration test weights, then press

enter

to

save the value and go to span calibration
(

WSPAN

).

5. Place test weights on scale. Press

down

again

to show the current WSPAN value, then press
the

Calibrate

softkey to calibrate span. When

complete, the new A/D count for the span
calibration is displayed. Press

enter

again to

save the span calibration value and go to the
next prompt (

WLIN

).

6. Five-point linearization (using the WLIN

parameter) provides increased scale accuracy

by calibrating the indicator at up to five
additional points between the zero and span
calibrations.

Linearization is optional: if you choose not to
perform linearization, skip the WLIN
parameter; if linearization values have
previously been entered, these values are reset
to zero during calibration. To perform
linearization, follow the procedure below:

With

WLIN

displayed, Press

down

to go to the

first linearization point (

POINT 1

). Press

down

again to show the weight value prompt (

WGT

1

), then

down

once more to show the weight

value. Place test weights on the scale, then use
the numeric keypad to enter the actual test
weight value. Press

enter

to save the value and

move to the calibration (

CAL 1

) prompt. Press

down

to show the current calibration value,

then press the

Calibrate

softkey to calibrate the

linearization point. When complete, the A/D
count for the linear calibration is displayed.
Press

enter

again to save the calibration value

and go to the next prompt (

POINT 2

).

Repeat for up to five linearization points. To
exit the linearization parameters, press the

up

key to return to WLIN.

7. The optional rezero function is used to

remove a calibration offset when hooks or
chains are used to hang the test weights.

NOTE: The rezero function cannot be used with
five-point linear calibration.

If no other apparatus was used to hang the
test weights during calibration, remove
the test weights and press

up

to return to

the CALIBR submenu.

If hooks or chains were used during
calibration, remove these and the test
weights from the scale. With all weight
removed, go to the REZERO parameter,
then press

down

to show the current zero

value. Press the

Calibrate

softkey to adjust

the zero and span calibration values. Press

enter

or

up

to return to the CALIBR

submenu.

8. Press

up

to return to the SCALES menu, or

press the

Save and Exit

softkey to exit setup

mode.

4.3

Serial Command Calibration

To calibrate the indicator using serial commands, the
indicator serial port must be connected to a terminal or
personal computer. See Section 2.3.3 on page 9 for
serial port pin assignments; see Section 9.0 on
page 84
for more information about using serial
commands.

07/14/2006

01:57PM

Calibrate

WZERO

WVAL

REZERO

1291

Deadload Calibration

+/-

Millivolts

Zero Count =>

Last Zero

Temp Zero

CALIBR

SCALE1

SCALES