Configuration

29

Secondary Units (SECNDR) and Tertiary Units (TERTIA) Parameters

DECPNT

888888.8
8888888
8888880
8888800
8.888888
88.88888
888.8888
8888.888
88888.88

Decimal point location. Determines the location of the decimal point or dummy zeros in the
secondary or tertiary units display.

DSPDIV

2D
5D
1D

Display divisions. Selects the value of minimum division size of the displayed weight for
secondary or tertiary units display.

UNITS

KG
G
OZ
TN
T
GN
TROYOZ
TROYLB
LT
CUSTOM
OFF
NONE
LB

Specifies secondary or tertiary units for displayed and printed weight. Values are:
LB=pound; KG=kilogram; G=gram; OZ=ounce; TN=short ton; T=metric ton; GN=grain;
TROYOZ=troy ounce; TROYLB=troy pound; LT=long ton.

MULT

0.453592
0.000001–
9999999

Multiplier. Specifies the conversion factor by which the primary units are multiplied by to
obtain the secondary or tertiary units. The default is 0.453592, which is the conversion
factor for changing pounds to kilograms. See Section 10.10 on page 117 for a list of
multipliers.

To shift between primary, secondary, and tertiary units, press the UNITS key.

Rate of Change (ROC) Units Parameters

For applications using the ROC function, the primary scale should be configured with a finer resolution than the rate of change
(ROC) units to prevent a stepwise appearance in the ROC display. The ROC display step size (weight increment between
displayed values) can be roughly calculated as follows:

(updates_per_ROC_UNIT) * (PRIMARY_resolution / ROC_resolution)

For example, with INTERVL=30; REFRESH =0.1; UNITS=MIN; PRIMARY resolution at 0.1 LB and ROC resolution at 1.0
(LB/MIN):

-

INTERVL * REFRESH = 30 * 0.1 = 3.0 seconds per update (ROC data is flushed every 3.0 seconds)

-

With UNITS = MIN, there are 20 ROC data updates per ROC time unit: 60 seconds / 3.0 seconds per update

-

The resolution ratio of PRIMARY to ROC units is 0.1 (0.1 / 1.0)

-

This configuration provides a step size in the ROC display of 2 LB (2 LB increments between displayed values):

20 * 0.1 = 2

Note that setting both PRIMARY and ROC display resolutions to 1 LB would have resulted in a step size of 20 LB.

SCALES Menu, FORMAT Submenu, SPLIT = OFF

Parameter

Choices

Description

Table 3-3. SCALES Menu, FORMAT Submenu Parameters, SPLIT = OFF (Continued)