ACCESS

ON
HIDE
OFF

Specifies the access allowed to setpoint parameters shown by pressing the
Setpoint softkey in normal mode.

ON: Values can be displayed and changed
HIDE: Values cannot be displayed or changed
OFF: Values can be displayed but not changed

NAME

NONE, 1–60

Specify the number of an assigned prompt. Up to 60 prompt names can be specified
on the PROMPTS submenu of the FEATURE menu.

SLOT

slot_number

Lists all available digital I/O slots. This parameter specifies the slot number of the
digital I/O card referenced by the DIGOUT parameter.

DIGOUT

bit_number

Lists all digital output bits available for the specified SLOT. This parameter is used to
specify the digital output bit associated with this setpoint. Use the DIG I/O menu to
assign bit function to OUTPUT.

For continuous setpoints, the digital output becomes active (low) when the condition
is met; for batch setpoints, the digital output is active until the setpoint condition is
met.

BRANCH

0
1-100

Specifies the setpoint number to which the batch sequence is to branch if the current
setpoint is not satisfied upon initial evaluation.

The special value zero indicates that no branch is taken.

TIME

time

For TOD setpoints, specifies the time at which the setpoint becomes active. The
format used to enter the time (12-hour or 24-hour) is based on the value specified for
the TIMEFMT parameter on the FEATURE menu.

DURATION

hh:mm:ss

For TOD setpoints, specifies the length of time that the digital output associated with
this setpoint changes state. The value is entered in hours, minutes, and seconds
(hh:mm:ss). All other operations associated with this setpoint (print, tare, or
accumulate) are performed at the end of the specified duration.

NSAMPLE

1–65535

For AVG setpoints, specify the number of A/D samples used to calculate the average
weight.

SOURCE

source_scale

Specify the scale number used as the source for the setpoint.

DIN SLOT

slot_number

For DIGIN and DINCNT setpoints, specify the slot number from which digital inputs
will be read.

DIN MASK

digital_input_mask

For DIGIN and DINCNT setpoints, specify the bits used as inputs to the setpoint. Use
the Select softkey to select bits.

VUNDER

0–9999999

For CHKWEI setpoints, specifies the lower weight limit.

VOVER

0–9999999

For CHKWEI setpoints, specifies the upper weight limit.

DUNDER

digital_output

For CHKWEI setpoints, specifies the bit number of digital output activated when the
scale weight is less than the VUNDER value specified.

DACCEPT

digital_output

For CHKWEI setpoints, specifies the bit number of digital output activated when the
scale weight between the VUNDER and VOVER values specified.

DOVER

digital_output

For CHKWEI setpoints, specifies the bit number of digital output activated when the
scale weight is greater than the VOVER value specified.

COAST

0–65535

For PLSCNT setpoints, specifies the time delay (in 0.1-second intervals) inserted
between reaching the setpoint target value and capture of the actual pulse count.

SENSE

NORMAL
INVERT

Specifies whether the value of the digital output associated with this setpoint is
inverted when the setpoint is satisfied.

SETPTS Menu

Parameter

Choices

Description

Table 8-2. Setpoint Menu Parameters (Continued)