Rice Lake 920i Installation Manual V4.0 User Manual

Page 61

Advertising
background image

Calibration

55

4. Remove all weights from the scale. If chains

or hooks are used, place them on the scale.
Press the

Calibrate Zero

button to perform the

zero calibration (see Figure 4-6). A message
box appears when the process is complete.

Figure 4-6.

iRev 4 Zero Calibration Display

5. Apply the test weights to the scale. Press the

Calibrate Span

button to perform the span

calibration (see Figure 4-7). A message box
appears when the process is complete.

Figure 4-7.

iRev 4 Span Calibration Display

6. If the option for chains or hooks was selected

in step 3, the Rezero display is shown (see
Figure 4-8). Remove all weights from the
scale, including chains or hooks. Press the

Re-Zero

button to calibrate the zero offset.

Figure 4-8.

iRev 4 Rezero Display

7. If performing a multi-point linear calibration,

up to five other calibration weight values can
be entered on the display shown in Figure 4-9.
The weights must be in ascending order and
must not include zero or the span weight.
Enter the weight values and click the

Go

button to calibrate each point.

Figure 4-9.

iRev 4 Linear Calibration Display

8. Review the new calibration values then click

Finish

to close the Calibration Wizard. To

restore the current calibration values, click

Cancel

.

Figure 4-10.

iRev 4 Calibration Values Display