2 scale capacity, 3 load cell mv, 4 total load cell build – Rice Lake BCi Belt Scale - Installation & Operation Manual Version 2.00 User Manual

Page 46: 5 rate unit time, 6 filter (s), 7 filter threshold (divisions), 8 speed unit time, 9 fixed speed, 10 unit of measure, 11 unit of rate

Advertising
background image

40

BCi

Installation & Operation Manual

There are many parameters in the
Supervisor Mode (shown in the following
sections), but most will not be used for a
basic installation. The parameters that

must be set prior to calibration include the following:

Calibration Weight (if using static weights to calibrate)

Idler Spacing

Number of Idlers

Belt Length

Load Cell MV

Total Load Cell Build

The other parameters that are commonly set are the
filtering values but they can be done after setup and
calibration.

The angle of inclination is not required with
Rice Lake Weighing Systems frames due to
laterally opposed shearbeams.

4.2

Scale Capacity

This parameter defines the maximum rated capacity
for the belt scale or the maximum amount of weight
that can pass along the belt scale per hour.

The default parameter is measured in tons/hour.

Enter the maximum rate in tons using the numeric
keypad and press the

enter

key to save that parameter.

4.3

Load Cell MV

Enter the average mv/v rating of all the load cells in
the system.

Use the numeric keypad to enter that value which
must be greater than 0. Press the

enter

key to save that

parameter.

4.4

Total Load Cell Build

This parameter is the total load cell capacity of the all
the load cells in system. If you have eight load cells
for example and each one has a rated capacity of XX,
take that capacity and times it by 8 to get the total load
cell capacity to be entered into the BCi integrator.

4.5

Rate Unit Time

This parameter defines what unit of time the rate will
be displayed in on the BCi. Available selections are:

Hr. - hour (default)

Min - minute

Sec - seconds

Press the

enter

key to toggle between the listed

choices above. The default rate unit time is Hr.

4.6

Filter (s)

This parameter defines the number of seconds
filtering samples to average the scale load. It is
measured in tenths of seconds.

4.7

Filter Threshold (divisions)

This is the number os division the load must maintain
for the filter to be enabled. If the load sees a change in
weight that is greater than this value, the filter is
flushed.

4.8

Speed Unit Time

This parameter defines how the belt speed is displayed
on the BCi. The selections are displayed in the
following units.

Hr. - hour

Min - minute (default)

Sec - seconds

Press the

enter

key to toggle between the listed

choices above. The default speed unit time is
displayed in minutes.

4.9

Fixed Speed

This parameter allows a technician to configure a
fixed belt speed for the BCi. The BCi will no longer
calculate speed using the speed sensor.

The default value is 0 and if a value is set then the BCi
will used the fixed value. The unit of measure is how
the BCi has been configured for speed so usually it
will be either ft/min or m/min.

The user can change the time units to be either
seconds or hours. Enter a numeric value and press the

enter

key to save that value.

4.10 Unit of Measure

The unit of measure parameter defines how the belt is
measured and speed calculated. The selections are
displayed in the following units.

Ft - feet (default)

M - meters

Press the

enter

key to toggle between the listed

choices above. The default speed is shown as Ft/Min.

4.11 Unit of Rate

The unit of rate parameter defines how the rate is
displayed. The selections are displayed in the
following units:

TN (tons) - (default)

lb (pounds)

kg (kilograms

Note

Note