0 calibration, 1 speed sensor calibration, Calibration – Rice Lake BCi Belt Scale - Installation & Operation Manual Version 2.00 User Manual

Page 52

Advertising
background image

46

BCi

Installation & Operation Manual

5.0

Calibration

There are two components of the Belt Scale Conveyor system need to be calibrated for the system to work. Those
component parts include:

Speed sensor

Integrator (BCi)

The speed sensor calibration must be done prior to the BCi calibration.

5.1

Speed Sensor Calibration

A belt conveyor scale shall be equipped with a belt speed or travel sensor that will accurately sense the belt speed
or travel whether the belt is empty or loaded.

Use the following steps to calibrate the speed sensor.

1. Select the

Belt Calibration

softkey from the Supervisors menu.

2. The integrator will change the softkeys to

Start

,

Stop

, and

Exit

.

3. A reference point needs to be marked on the conveyor belt and a reference needs to be marked on the

conveyor frame. This will give the operator a reference to count the number of revolutions the belt
travels during the speed calibration. The more revolutions in a test, the better the speed and distance
accuracy.

4. Press the

Start

softkey. This will cause the integrator to change screens displayed the number of pulses

counted, the frequency of the pulses and the time the test is running as shown below.

Figure 5-1. Pulses Counted

The screen illustrates a live display of how the belt is working. If the pulses and frequency are not
changing then there is a problem with the speed sensor and can be a sign that either the wiring or the
sensor itself is bad. During this cycle of the calibration procedure, the operator needs to count the
number of belt revolutions.

5. Pressing the

Stop

softkey will stop the speed sensor calibration. The integrator will then prompt the user

Note

Exit =>

SCALE #1

Start

11/23/2011

08:13AM

96489 Pulse

261 Per Sec

889.25

Time Running

Stop