Material Calibration Mode

Use this calibration mode if you want to calibrate the scale with a known amount of material. The material must
be pre-weighed or post weighed.

Use the following steps to perform a material calibration.

1. Press the

Material

softkey on the BCi display. The BCi will display

Span Cal

and

Exit =>

softkeys.

2. The

Span Cal

softkey will initiate the calibration sequence for the span calibration. This is similar to the

zero cal but material will be passed over the scale during the calibration test.

3. The operator will press

Start

, the integrator will start taking span averages. At some point the operator

will then press the

Finish

softkey to end the sequence.

4. The BCi will prompt the operator for the amount of material that was weighed. The operator may key in

the new value and press

enter

or cancel to exit with no changes.

Figure 5-6. Accept New Span

5. The process can be aborted at any time by pressing the

Exit =>

softkey.

Zero Cal Calibration Mode

By selecting this mode will calibrate the integrator based on a zero test. A zero calibration is based on the number
of belt revolutions is established during the belt speed sensor calibration.

Use the following steps to perform a Zero Cal calibration.

1. Press the

ZERO

key on the integrator. The previous zero information will be displayed with the zero scale

menu.

2. The operator will press the Start key.

3. The Zero Cal test will run, and after the test is completed the new zero error% will be shown on the

display. The operator needs to press either the Yes softkey to accept that value or No softkey to not see
the new zero.

4. There is also a config mode zero test which goes through the same steps as the zero cal calibration but

this is accessed through the Integrator calibration softkey on the main menu.

5.3

Complete System Calibration Test Used In Conjunction with integrator
Calibration

There are three types of tests that are also used in conjunction with calibrating the complete system. They are:

Material testing

Simulated (resistor) load testing.

Maintenance testing

Material testing is used only with the material calibration and the simulated (resistor) load testing is used only
with the Auto Calibration and the Timed Cal.

The following sections describe how to perform a material test and a simulated load test.

11/23/2011

08:13AM

No

Test Time

Test Accumulator

Old Span Error %

Yes

Span Error %

13.50

147.4

3274

25.00

Accept New Span