3. PRECAUTIONS

To get the optimum performance from the balance and acquire accurate weighing data, note the

following:

3-1 Before Use

Install the balance in an environment where the

temperature and humidity are not excessive. The best
operating temperature is about 20°C / 68°F at about
50% relative humidity.

Install the balance where it is not exposed to direct

sunlight and it is not affected by heaters or air
conditioners.

Install the balance where it is free of dust.

Install the balance away from equipment which

produces magnetic fields.

Install the balance in a stable place avoiding vibration

and shock. Corners of rooms on the first floor are best,
as they are less prone to vibration.

The weighing table should be solid and free from

vibration, drafts and as level as possible.

Level the balance by adjusting the leveling feet and

confirm it using the bubble spirit level.

Ensure a stable power source when using the AC

adapter.

Warm up the balance for at least 30 minutes. Plug in

the AC adapter as usual.

Calibrate the balance before use or after having

moved it to another location.

Caution

Do not install the balance where flammable or corrosive gas is present.