22

7-2 Self Check Function

This function self-checks the internal performance of the balance.

Operation

1 Press and hold the MODE key until RESPONSE is

displayed, then release the key.

2 The balance automatically starts to check its internal

performance

3 After self-check, the balance displays the checking

result and returns to the weighing mode.

e.g.

OK

The example above indicates that the result of the self

check is good.

Note
If improper performance is found in the self check, the
balance displays
CH no. Contact the local A&D dealer
for repair.




















Press and hold

the key

Release the key

Displays the result.