Programmable-unit – Rice Lake GF Series, A&D Weighing User Manual

Page 55

Advertising
background image

53

14. PROGRAMMABLE-UNIT

This is a programmable unit conversion function. It multiplies the weighing data in grams by an
arbitrary coefficient set in the function table and displays the result.

The coefficient must be within the range between the minimum and maximum shown below. If the
coefficient set is beyond the range, an error is displayed and the balance returns to the coefficient
setting mode, prompting to enter an appropriate value. A coefficient of 1 was set at the factory.

Model

Minimum coefficient

Maximum coefficient

GF-200/300/400/600/800/1000 1000
GF-1200/2000/3000/4000/6100 100
GF-6000/8000

0.000001

10

Operation

1 Press and hold the SAMPLE key until ba5fnc of the function table is displayed.

2 Press the SAMPLE key several times to display mlt

Cadj

.

3 Press the PRINT key. The balance enters the mode to confirm or set the coefficient.

Confirming the coefficient

4 The current coefficient is displayed with the first digit blinking.

When it is not to be changed, press the CAL key and proceed to step 6.

When it is to be changed, press the RE-ZERO key

and proceed to step 5.

Setting the coefficient

5 Set the coefficient using the following keys.

SAMPLE key

To select a digit to change the

value. The selected digit blinks.

RE-ZERO key

To change the value.

MODE key

To change the decimal point

position.

Each time the switch is pressed,

the decimal point position
changes as follows:

PRINT

key

To store the new setting,

display end and go to step 6.

CAL key

To cancel the new setting and

go to step 6.

Quitting the operation

6 The balance displays Unit . Press the CAL key to exit the programmable-unit function and

return to the weighing mode.

Using the function

Press the MODE key to select the programmable-unit (no display on the unit section). Perform

weighing as described in “6-1 Basic Operation (Gram Mode)”. After weighing, the balance
displays the result (weighing data in grams x coefficient).

Advertising