Maintenance, Troubleshooting, 1 checking the balance performance and environment – Rice Lake GF Series, A&D Weighing User Manual

Page 70

Advertising
background image

68

19. MAINTENANCE

Do not use organic solvents to clean the balance. Clean the balance with a lint free cloth that

is moistened with warm water and a mild detergent.

Do not disassemble the balance. Contact the local A&D dealer if the balance needs service or

repair.

Use the original packing material for transportation.


20. TROUBLESHOOTING

20-1 Checking the Balance Performance and Environment

The balance is a precision instrument. When the operating environment or the operating method is

inadequate, correct weighing can not be performed. Place a sample on the pan and remove it, and

repeat this several times. If the balance seems to have a problem with repeatability or to perform

improperly, check as described below. If improper performance persists after checking, contact the

local A&D dealer for repair.

Checking that the balance performs properly

Check the balance performance using the self-check function as described in “7-2 Self Check

Function”.

An error display appears when a malfunction is found.

Check the balance repeatability using an external weight. Be sure to place the weight in the

center of the weighing pan.

Check the balance repeatability, linearity and calibrated value using external weights with a

known value.

Checking that the operating environment or weighing method is proper

Operating environment

Is the weighing table solid enough (especially for the GF-200/300/400/600/800/1000, the

balance with a minimum weighing value of 0.001 g)?

Is the balance level? Refer to “3-1 Before Use”.

Is the operating environment free from vibration and drafts? For the GF-200/300/400/600/800/

1000, has the breeze break been installed?

Is there a strong electrical or magnetic noise source such as a motor near the balance?

Weighing method

Does the weighing pan touch the breeze ring or anything? Is the weighing pan installed

correctly?

Is the RE-ZERO key pressed before placing a sample on the weighing pan?

Advertising