Rice Lake GF Series, A&D Weighing User Manual

Page 41

Advertising
background image

39

CSV format

5if type 5

Separates the data of A&D standard format and the unit by a comma (,).
Outputs the unit even when the data is overloaded.
When the ID number and data number are added, outputs the ID number, data number and

weighing data in this order and separates each item by a comma and treats all the items as one
group of data.

Note
To add the ID number and data number, the function settings must be changed.

10-7

Description of the Data Format Added to the Weighing Data

Data number dout d-no 1

This format outputs the data number just before the data is transmitted using the RS-232C interface.

This format consists of six characters excluding the terminator.
When CSV format (5if type 5) is selected, the period (

. ) is replaced with a comma (

,

).


Note

The data number is added only to the weighing data that is stored in memory.

ID number

dout 5-id 1

The number to identify a specific balance.

This format consists of seven characters excluding the terminator.

Note

When the data described above is added to the weighing data, the output is in the
following order: ID number, Data number, and Weighing data.