5 description of the item “data output mode – Rice Lake GF Series, A&D Weighing User Manual

Page 38

Advertising
background image

36

10-5 Description of the Item “Data output mode”

The parameter setting of the “Data output mode ( prt )” applies to the performance when the “Data
memory ( data )” parameter is set to “2” (to store the weighing data) and when the data is
transmitted using the RS-232C interface.

Key mode

When the PRINT key is pressed with the stabilization indictor turned on, the balance outputs or
stores the weighing data and the display blinks one time.

Required setting

dout

prt 0 Key

mode

Auto print modes A and B

When the displayed value is stable and the conditions of “Auto print polarity”, “Auto print difference”
and reference value are met, the balance outputs or stores the weighing data.
When the PRINT key is pressed with the stabilization indictor turned on, the balance outputs or
stores the data and the display blinks one time.

Mode A: Required setting

dout prt

1

Auto print mode A (reference = zero)

dout ap-p

Auto print polarity

dout ap-b

Auto print difference

Example

“For weighing the added sample each time a sample is added, with
“ar-d” set to “1” (to adjust zero after the data is output).”

Mode B: Required setting

dout prt

2

Auto print mode B (reference =

last

stable

value)

dout ap-p

Auto print polarity

dout ap-b

Auto print difference

Example

“For weighing while a sample is added.”

Stream mode

The balance outputs the weighing data continuously regardless of the display condition. The display
does not blink in this mode. This mode is not available and the interval memory mode is used when
the “Data memory ( data )” parameter is set to “2” (to store the weighing data).

Required setting

dout prt

3 Stream

mode

dout data

0

Data memory function is not used.

ba5fnc 5pd

Display refresh rate

5if bp5 Baud

rate

Example

“For monitoring data on a computer”

Caution

The balance may not transmit the data completely at the specified refresh rate, depending on
the baud rate or data added to the weighing data such as time, date and ID number.