2 error codes – Rice Lake GF Series, A&D Weighing User Manual

Page 71

Advertising
background image

69

Is the sample placed in the center of the weighing pan?

Has the balance been calibrated using an external mass?

Has the balance been warmed up for 30 minutes before weighing?


Sample and container

Has the sample absorbed or lost moisture due to the ambient conditions such as temperature

and humidity?

Has the temperature of the container been allowed to equalize to the ambient temperature?

Refer to “3-2 During Use”.

Is the sample charged with static electricity? Refer to “3-2 During Use”.

The GF-200/300/400/600/800/1000 is prone to be charged with static electricity when the

relative humidity is low.

Is the sample of magnetic material such as iron? Caution is required for weighing magnetic

materials. Refer to “3-2 During Use”

20-2 Error Codes

Display Error code

Description

EC, E11

Stability error

The balance can not stabilize due to an environmental
problem. Prevent vibration, drafts, temperature
changes, static electricity and magnetic fields.

Refer to “3. PRECAUTIONS” for details on the
operating environment and “7. RESPONSE
ADJUSTMENT” about adapting the balance to the
environment.

To return to the weighing mode, press the CAL key.

Out of range error

The value entered is beyond the settable range.

Re-enter the value.

EC, E20

Calibration weight error

The calibration weight is too heavy. Confirm the
calibration weight value.

Press the CAL key to return to the weighing mode.

EC, E21

Calibration weight error

The calibration weight is too light. Confirm the
calibration weight value.

Press the CAL key to return to the weighing mode.

Overload error

A sample beyond the balance weighing capacity has
been placed on the pan.

Remove the sample from the pan.

Advertising