2 calibration using an external weight – Rice Lake GF Series, A&D Weighing User Manual

Page 26

Advertising
background image

24

8-2 Calibration Using an External Weight

This function calibrates the balance using an external weight.

Operation

1 Connect the AC adapter and warm up the balance

for at least 30 minutes with nothing on the pan.

2 Press and hold the CAL key until Calout is

displayed, and then release the key.

3 The balance displays Cal 0 .

If you want to change the calibration weight (a

list of usable weights is shown on page 23),
press the SAMPLE key and proceed to step 4.

If you use the calibration weight value stored in

the balance, proceed to step 5.

4 Specify the calibration weight value as follows:

SAMPLE

key To switch the display

condition to: “All of the
segments blinking”
(calibration weight selection
mode) or “The last two digits
blinking” (value adjustment
mode).

RE-ZERO

key To select the calibration

weight or adjust the value. In
the value adjustment mode,
-15 digits appear after +15
digits.

PRINT

key

To store the new weight

value. Even if the AC
adapter is removed, the data
is maintained in non-volatile
memory.

CAL key

To cancel the operation and

return to Cal 0 .

Note
Digit, when used for the GF series balance,
indicates a unit of minimum weighing value.



Select a

weight

To next page

Press and

hold the key

Release

the key

e.g. Calibration

weight value
1000.12 g

Advertising