37

Interval memory mode

The weighing data is periodically stored in memory.

Required setting

dout prt

3

Interval memory mode

dout data

2

Data memory function is used.

Stores weighing data.

dout int Interval

time

Example

“For periodical weighing without a computer command and to output
all of the data, to a computer, at one time”

10-6 Description of the Item “Data format”

A&D standard format

5if type 0

This format is used when the peripheral equipment can receive the A&D format. If an AD-8121B is
used, set the printer to MODE 1 or 2.

z

This format consists of fifteen characters excluding the terminator.

z

A header of two characters indicates the balance condition.

z

The polarity sign is placed before the data with the leading zeros. If the data is zero, the plus

sign is used.

z

The unit, consisting of three characters, follows the data.

DP (Dump print) format

5if type 1

This format is used when the peripheral equipment can not receive the A&D format. If an AD-8121B
is used, set the printer to MODE 3.

z

This format consists of sixteen characters excluding the terminator.

z

A header of two characters indicates the balance condition. No overload header is used.

z

The polarity sign is placed before the data, with spaces in place of leading zeros, if the data is

not zero or overloaded.

z

The unit, consisting of three characters, follows the data.

Unit

Data

Header

S T , + 0 0 0 1 2 . 7 8

1 1

g

CR LF

Terminator