The following table lists the various display messages and sequence when setting up the scale.

NOTE:

<-> means that you can toggle between two values.

Step

Function

Display

Available Parameters

1

With the scale turned
off, remove the
configuration jumper
as shown in
Figure 5-1.

Allows the scale to enter into either
configuration or calibration mode.

2

With the scale turned
off, simultaneously
press the On-Off/
Zero and Kg-Lb key.

Enters into
programming mode

StArt

Scale automatically advances to Step 3.

3

Indentifies the
software ID

IdEnt<->11007

Press the REWEIGH key to advance to
the next step.

4

Identifies software
version

Id<->11387

Press the REWEIGH key to advance to
the next step.

5

Allows selection of
decimal point setting

dOt<->000.0
Default = 0.0

To change the position of the decimal
point, press the Kg-Lb key to toggle
through the various options. To advance
to the next step, press the REWEIGH key.

6

This indicates the
maximum allowed
weight. Any weight
above this value will
cause StOP to appear
on the display while in
the weighing mode.

FULL<->XX.XXX
Default= 600 Lb

Note: This value is
dependent on the model
you have. Refer to the
serial number label on
your scale and set
appropriately.

Use the numeric data entry (See “Numeric
Data Entry” on page 12.
) to change the
value. To advance to the next step, press
the REWEIGH key.

7

This limit defines the
start of the weighing
threshold to
compensate for a key
press during the Tare
function.

Limit <->0.0

Use the numeric data entry (See “Numeric
Data Entry” on page 12.
) to see the
optimum weight value of 5 lbs. To
advance to the next step, press the
REWEIGH key.

8

Display divisions

rOUnd<->XXXX
Default = 0.2 Lb

To change the display divisions, press the
Kg-Lb key to toggle through the various
options.
0.2, 0.5, 1.0, 2.0, 5.0, 10.0, 20.0.
To advance to the next step, press the
REWEIGH key.

9

Double ranges limit

drAnGe <-> 0

This parameter is preset from the factory.
To advance to the next step, press the
REWEIGH key.

10

Weight algorithm initial
tolerance

A tOL <-> 10

This parameter is preset from the factory.
To advance to the next step, press the
REWEIGH key.

11

Weight algorithm initial
exponent

A LEn <-> 8

This parameter is preset from the factory.
To advance to the next step, press the
REWEIGH key.

12

Weight algorithm
maximal exponent

A t INE <-> 10

This parameter is preset from the factory.
To advance to the next step, press the
REWEIGH key.

Table 5-1. Configuration Mode Menu