Reset to factory defaults – Rice Lake Portable Bariatric Wheelchair Scale Single Ramp (350-5) - Rice Lake Bariatric Wheelchair Scale User Manual

Page 20

Advertising
background image

16

Bariatric Wheelchair Technical and Operating Manual

Reset To Factory Defaults

The configuration parameters may be reset to factory defaults while in the configuration mode. To do so, enter

the configuration mode. Press the

REWEIGH

key once to advance to displaying the software version (ID <->

11305). Press and hold the

BMI

key until

=DEF=

is displayed. At this point, if you wish to perform the default

function, press the

REWEIGH

key and the scale will show

DONE

, then it will restart. If you do not want to perform

the default function, press the

ON/OFF

key.

After performing the default function, the scale will require re-calibration. In addition, the scale will be set

up to a capacity of 600 lbs. The capacity varies by model - refer to the serial label on your scale and

reconfigure the capacity (FULL) and display divisions (rOUnd) as necessary in the configuration menu.

17

Determines the
automatic shut off
time when the scale is
not in use. Options are
between one and 20
minutes.
This is used when the
unit is battery
operated.

t-OFF <-> 5

Press and hold the Kg-Lb key to scroll
through the furthest right hand digit on the
display (0-9). If you want a value from 10
to 19, with the display sitting at 1, double
press the Kg-Lb key and 10 appears.
Press and hold the Kg-Lb key to begin
scrolling 10-19. If you want a value of 20,
with the display sitting at 2, double press
the Kg-Lb key and 20 will appear.
To advance to the next step, press the
REWEIGH key.

18

StArt or SAvE

If no changes were made in the
parameters the display will automatically
show StArt and then returns to normal
weighing mode.
If a parameter was changed, the display
will show SAvE. To save changes made,
press the REWEIGH key. To return to
weigh mode without saving changes,
press the Kg-Lb key.

19

Con En

Replace the configuration jumper.

Step

Function

Display

Available Parameters

Table 5-1. Configuration Mode Menu