Bariatric Wheelchair Technical and Operating Manual - Scale Calibration

17

6.0

Scale Calibration

Before you can calibrate the scale, verify and set all scale parameters which are noted in “Scale Configuration”

on page 11.
Remove the configuration/calibration jumper (see Figure 5-1 on page 11). Turn the scale on. While

StArt

is

displayed, press and hold both the

Kg-Lb

key and the

REWEIGH

keys until

CAL

is displayed.

To do a calibration on the scale, the scale must be turned off.

NOTE: The calibration weight must be no less than 60 lb (28 kg) and no more than 300 lb (135 kg).

The following chart illustrates the calibration procedure.

Table 6-1. Calibration Menu

Step

Function

Display

Available Parameters

1

Turn the scale on.
While START is
displayed, press and
hold the Kg-Lb and
REWEIGH keys until
CAL is displayed.

Enters into calibration
mode

StArt

The scale automatically advances to
Step 2.

2

Cal mode entered

CAL

To advance to the next step, press the
REWEIGH key.

3

Sets the value of the
calibration weight
you are going to use
for calibrating the
span value of the
scale.

LOAd <-> XXX.X

Use the numeric data entry (explained in
“Numeric Data Entry” on page 12) to set
a calibration weight.
To advance to the next step, press the
REWEIGH key.

4

CLEAr

Clear the platform and be sure of the
scale’s stability.
To advance to the next step, press the
REWEIGH key.

5

Calibrate Zero

- - - - -
PUT<->XXXX

Place the requested weight on the scale.
This will display for a few seconds.
To advance to the next step, press the
REWEIGH key.

6

Calibrate Span

CAL
FAC tOr <-> X.XXX

This will be displayed for a few seconds
and shows the current calibration factor.
To advance to the next step, press the
REWEIGH key.

7

SAvE

To save the new calibration, press the
REWEIGH key.
To exit without saving the calibration
press the ZERO key.

8

dONE

The scale displays that it has saved that
calibration value and automatically
advances to the last step.

9

StArt

The scale reboots.

10

Con En

Replace the configuration jumper.