Bariatric Wheelchair Technical and Operating Manual - Indicator Setup

5

3.0

Indicator Setup

To protect the indicator during transit, the indicator is shipped in a separate box inside the main shipping

container. The following sections apply to indicator connections and setup.

3.1

Load Cell Connections

Your scale comes with ten feet of load cell cable which comes from the bottom of the scale (shown in

Figure 3-1). The load cell cable must be connected to the indicator display.
To gain access to the load cell connection point, do the following:

1. Unscrew and remove the tilt stand bracket from the indicator.

Unscrew tilt stand knobs
and remove tilt stand from
indicator

Figure 3-1. Remove Tilt Stand From Indicator

2. Remove the four back retaining screws as shown in Figure 3-2 and remove the back cover to the

indicator.

Figure 3-2. Remove Back Cover of Indicator