Bariatric Wheelchair Technical and Operating Manual - Troubleshooting and Testing

23

8.0

Troubleshooting and Testing

The following sections describe basic service and maintenance procedures.

Problem

Symptom Description

Solution

Scale does not turn
on

Dead battery

Connect the scale to a power source.

Faulty electrical outlet

Use a different electrical outlet.

Bad power supply

Replace adaptor.

Questionable weight
or the scale does not
zero

External object is interfering with the scale Remove the interfering object from the scale.

Display did not show 0.0 before weighing

Help the patient off the scale, zero the scale
and begin the weighing process again.

Scale is not placed on a level floor

Ensure the scale is level and begin the
weighing process again.

Scale is out of calibration

Check the weight with a known weight value.

Improper tare

Place the tare item on the scale. Press
REWEIGH. Once the weight of the item is
displayed, press TARE. Place the patient
back on the scale. Press the REWEIGH
button again.

Consistently high or
low weights

Indicator not properly adjusted to zero

Zero the indicator according to the indicator
manual

Platform is binding

Obtain adequate clearance for free platform
movement

Indicator is not calibrated

Calibrate accordingv to the indicator manual

Load cells faulty

Test and replace load cells if necessary

Feet touching deck underside

Adjust feet downward to provide clearance

Weighing is
performed but the
display shows
WEIGH and
REWEIGH every few
seconds; the
weighing process
takes too long and no
weight is displayed

The patient is not sitting still

Ask the patient to be still

Erratic weight
readings

Vibration near the scale

Remove source of the vibration or move
scale

Platform not level to within 1/4 inch

Level the scale by adjusting the feet or
shimming if necessary

Load cell or cable water damage

Replace

Debris under load cells or platform

Clean

indicator faulty

Use simulator or test indicator for stability.
Service indicator

Display stays at zero

Indicator faulty

Service indicator

Load cell connections are faulty

Check the cable connections in junction box
and at the indicator

The display shows a
STOP message

The load on the scale exceeds the
capacity of the scale

Remove the excess weight and use the scale
according to manufacture’s specs.

Table 8-1. Troubleshooting Table