Bariatric Wheelchair Technical and Operating Manual - Scale Configuration

15

Option 6 allows you to
either enable or
disable the hold
function on the scale

OP6 = 1

Hold the Kg-Lb key for browsing through
the options. Double click on the Kg-Lb
key to change the options value or
to advance to the next step, press the
REWEIGH key.
0=Disable
1=Enable

Option 7 allows you to
either enable or
disable the baby scale
function.

OP7=0

Hold the Kg-Lb key for browsing through
the options. Double click on the Kg-Lb
key to change the options value or
to advance to the next step, press the
REWEIGH key.
0=Disable
1=Enable
Note: Leave set to 0

Option 8 allows you to
choose between 9
volts and 6 volts

OP8 = 1

Hold the Kg-Lb key for browsing through
the options. Double click on the Kg-Lb
key to change the options value or
to advance to the next step, press the
REWEIGH key.
0=9 volts
1=6 volts
Note: Leave set to 1

Option 9 allows you to
select the
communications
protocol.

OP9 = 1

Hold the Kg-Lb key for browsing through
the options. Double click on the Kg-Lb
key to change the options value or
to advance to the next step, press the
REWEIGH key.
0= escape
1= standard protocol

Option 10 allows you
to enable the user
menu.

OP10 = 1

Hold the Kg-Lb key for browsing through
the options. Double click on the Kg-Lb
key to change the options value or
to advance to the next step, press the
REWEIGH key.
0= enable
1= disable

16

Set Option defaults

Set < - > 1

To change, press the Kg-Lb key.
0 - default
Options = (OPx)
1 = US default (US defaults are OP1-1,
OP2-0, OP3-0, OP4-0, OP5-0, OP6-1,
OP-7-0, OP8-1, OP9-1, OP10-1).
2 = European 1 default
3 = European 2 default
9 = Used if OP1 through OP10 have been
changed to something other than the
factory default settings.

Note: Always set to 1 or 9 for US
indicators. To advance to the next step,
press the
REWEIGH key.

Step

Function

Display

Available Parameters