1 test mode – Rice Lake Portable Bariatric Wheelchair Scale Single Ramp (350-5) - Rice Lake Bariatric Wheelchair Scale User Manual

Page 28

Advertising
background image

24

Bariatric Wheelchair Technical and Operating Manual

8.1

Test Mode

The test mode menu is a special mode used for checking four very important parameters which are useful in

knowing the system’s state and for troubleshooting. Entry into this mode is possible only when the scale is turned

off. Turn the scale on. While

Start

is displayed, press and hold the

REWEIGH

key until both the middle arrows

appear on the display.
The test mode has four parameters. They are:

Weighing

Internal count

Battery indication

Calibration factor

Alternating between the parameters is performed by pressing the

REWEIGH

key.

Press the

Kg-Lb

key to zero the scale in test mode.

Press the

Kg-Lb

+

REWEIGH

to exit weighing mode.

Power On-Off

START

REWEIGH Key

TEST

Version

BAT

A-D

FACTOR

At

Press and
hold

Power Off

Remove Config/
Calibr. Jumper

Figure 8-1. Test Mode Menu Structure

Table 8-2 lists the various display messages when testing the scale.

The display shows
LO Bat message

The battery is low

Recharge the battery.

The display shows
Err message as
detailed in the table
below

Err 2

Low saturation state (low A/D)

The load cell is not connected properly.
Check the cables and mechanical
connections. If the problem persists, replace
the set of load cells.

Err 3

High saturation state (high A/D)

See Err 2

Err 6

Unstable weight. Cannot calibrate

Check the load cells’ mechanical
surroundings and see that nothing touches
them and that the cables are properly
welded.

Err 7

Mathematical error; division by zero.
Cannot calculate calibration factor.

This error will show when trying to calibrate
the unit with no calibration weight on the unit.

Problem

Symptom Description

Solution

Table 8-1. Troubleshooting Table