2 replacing the summing board – Rice Lake Portable Bariatric Wheelchair Scale Single Ramp (350-5) - Rice Lake Bariatric Wheelchair Scale User Manual

Page 32

Advertising
background image

28

Bariatric Wheelchair Technical and Operating Manual

9.2

Replacing the Summing Board

The signal trim summing board is easily accessible through an access plate located on the side of the Rice Lake

Bariatric Wheelchair Scale. Use the following steps to wire up the summing board.

1. Remove the two #10 x 3/8" screws.

Figure 9-5. Remove the Screws From the Access Plate

2. Slide the summing board out of the scale deck.
3. Connect the wires to the indicator terminal as shown in Table 9-2.

Table 9-2. Junction Box Connections

Cable Color Code

Junction Box

Red

+ Excitation

Black

- Excitation

Green

+ Signal

White

- Signal

Bare

Shield

I n d i c a t o r
Te r m i n a l
Location

Figure 9-6. Summing Board

4. When the interface cable is protected and in its final position, complete connections to the indicator.
5. Check all strain relief fittings for tightness.
6. Slide the summing board back into the floor scale cutout.
7. Secure the Rice Lake Bariatric Wheelchair Scale side plate with the two screws.