Bariatric Wheelchair Technical and Operating Manual - Maintenance

27

Figure 9-3. Foot Pad - Side View

Route the load cell cables near each corner so that the

cable is free from possible contact with each foot.

Hold the cable in position with the supplied

adhesive-backed cable ties.
Do not cut load cell cables. Coil extra cable before it

enters the junction box, tie with cable ties, and insert

the coils into the channel near the junction box.
After coiling excess cable, pass each individual end of

load cell cable through its grommet in the junction

box cover (or through cable fittings in the NEMA 4X

junction box).
Corner correction trimming and calibration is

necessary after load cell replacement. Follow

instruction in Section 6.0 on page 17.

9.1.1

Load Cell Wiring to Junction Box

The four load cells are each wired to their respective

terminals in the junction box according to the corner

numbering system shown in Figure 9-4 and the

coloring code in Table 9-1.
Pull excess cable out of the junction box enclosure

and tighten the cable grips with a wrench. To be

watertight, the cable grips must be tightened to the

point where the rubber sleeving begins to protrude out

of the hub. Finally, pull on each of the four cables to

make sure that they do not slip.

1

4

2

3

Figure 9-4. Corner Numbering - Top View

Table 9-1. Load Cell Wiring

Cable Color Code

J-Box Terminal

Red

+Excitation

Black

-Excitation

Green

+Signal

White

-Signal