SURVIVOR 390HE Installation Manual

3.0

Configuration

To configure the 390HE indicator, the indicator must

be placed in setup mode. The setup switch is accessed

by opening the enclosure cover and pressing the setup

switch mounted on the end of the pushbutton switch

board, next to the

ZERO

button. Press the switch once

to enter setup mode, then close the cover and tighten

the cover screws.
When the indicator is placed in setup mode, the word

CONFIG

is shown on the display. The CONFIG menu

is the first of seven main menus used to configure the

indicator. Detailed descriptions of these menus are

given in Section 3.2. When configuration is complete,

return to the CONFIG menu and press the (

ZERO

)

key to exit setup mode.

3.1

Configuration Methods

The 390HE indicator can be configured by using the

front panel keys to navigate through a series of

configuration menus or by sending commands or

configuration data to the EDP port. Configuration

using the menus is described in Section 3.1.3.
Configuration using the EDP port can be accomplished

using either the EDP command set described in
Section 5.0 on page 22 or the Revolution

configuration

utility.

3.1.1

Revolution Configuration

The Revolution configuration utility provides the

preferred method for configuring the 390HE indicator.

Revolution runs on a personal computer to set

configuration parameters for the indicator. When

Revolution configuration is complete, configuration

data is downloaded to the indicator.

Figure 3-1. Sample Revolution Format Display

Revolution supports both uploading and downloading

of indicator configuration data. This capability allows

configuration data to be retrieved from one indicator,

edited, then downloaded to another.
To use Revolution, do the following:

1. Install Revolution on an IBM-compatible

personal computer running Windows

®

3.11 or

Windows 95. Minimum system requirements

are 8MB of extended memory and at least

5MB of available hard disk space.

2. With both indicator and PC powered off,

connect the PC serial port to the RS-232 pins

on the indicator EDP port.

3. Power up the PC and the indicator. Use the

setup switch to place the indicator in setup

mode.

4. Start the Revolution program.

Figure 3-1 shows an example of one of the Revolution

configuration displays.
Revolution provides online help for each of its

configuration displays. Parameter descriptions

provided in this manual for front panel configuration

can also be used when configuring the indicator using

Revolution: the interface is different, but the

parameters set are the same.

3.1.2

EDP Command Configuration

The EDP command set can be used to configure the

390HE indicator using a personal computer, terminal,

or remote keyboard. Like Revolution, EDP command

configuration sends commands to the indicator EDP

port; unlike Revolution, EDP commands can be sent

using any external device capable of sending ASCII

characters over a serial connection.
EDP commands duplicate the functions available

using the indicator front panel and provide some

functions not otherwise available. EDP commands can

be used to simulate pressing front panel keys, to

configure the indicator, or to dump lists of parameter

settings. See Section 5.0 on page 22 for more

information about using the EDP command set.