3 setting the digital filter parameters, 7 test mode, Section 8.7 on – Rice Lake Survivor 390HE Hostile Environment Digital Indicator User Manual

Page 40: Caution

Advertising
background image

36

SURVIVOR 390HE Installation Manual

8.6.3

Setting the Digital Filter Parameters

Fine-tuning the digital filter parameters greatly

improves indicator performance in heavy-vibration

environments. Use the following procedure to

determine vibration effects on the scale and optimize

the digital filtering configuration.

1. In setup mode, set all three digital filters

(DIGFL1, DIGFL2, DIGFL3) to 1. Set

DFTHRH to NONE. Return indicator to

normal mode.

2. Remove all weight from the scale, then watch

the indicator display to determine the

magnitude of vibration effects on the scale.

Record the weight below which all but a few

readings fall. This value is used to calculate

the DFTHRH parameter value in Step 4.
For example, if a heavy-capacity scale

produces vibration-related readings of up to

50 lb, with occasional spikes to 75 lb, record

50 lb as the threshold weight value.

3. Place the indicator in setup mode and set the

digital filters (DIGFLx) to eliminate the

vibration effects on the scale. (Leave

DFTHRH set to NONE.) Reconfigure as

necessary to find the lowest effective values

for the DIGFLx parameters.

4. With optimum values assigned to the DIGFLx

parameters, calculate the DFTHRH parameter

value by converting the weight value recorded

in Step 2 to display divisions:

threshold_weight_value / DSPDIV

In the example in Step 2, with a threshold

weight value of 50 lb and a display division

value of 5D:

50 / 5D = 10.

DFTHRH should be

set to 10DD for this example.

5. Finally, set the DFSENS parameter high

enough to ignore transient peaks. Longer

transients (typically caused by lower vibration

frequencies) will cause more consecutive

out-of-band readings, so DFSENS should be

set higher to counter low frequency transients.
Reconfigure as necessary to find the lowest

effective value for the DFSENS parameter.

8.7

Test Mode

In addition to normal and setup modes, test mode

provides a number of diagnostic functions for the

390HE, including:

Display raw A/D count

Set A/D offset and gain calibration

Reset configuration parameters to default

values

Print indicator configuration

To enter test mode, press and hold the setup switch

until the front panel display shows the word

TEST

.

After about three seconds, the test mode display

automatically shifts to the first test menu function,

A/DTST.

A/D calibration functions, ADOFFS and
ADGAIN, must be used only by qualified
service personnel, and only after

replacing A/D converter components. Improper A/D
calibration may render the indicator unusable.

Figure 8-5 shows the Test Menu structure; Figure 8-6

shows the front panel key functions in test mode. Note

that, because the Test Menu functions are all on a

single menu level, the

GROSS/NET

( ) key has no

function. Press the

ZERO

( ) key to exit test mode.

Table 8-5 on page 37 summarizes the test menu

functions.

Figure 8-5. Test Menu

Caution

A/DTST

TEST

PRTCFG

DEFLT

Advertising