EDP Commands

23

5.1.2

Reporting Commands

Reporting commands (see Table 5-2) send specific

information to the EDP port. These commands can be

used in both setup mode and normal mode.

5.1.3

The RESETCONFIGURATION Command

The RESETCONFIGURATION command can be

used to restore all configuration parameters to their

default values. Before issuing this command, the

indicator must be placed in setup mode or test mode

(press and hold setup switch for approximately four

seconds to enter test mode).

Sending this command, followed by the RS (RESET)

EDP command, is equivalent to using the DEFLT

function on the TEST menu. See Section 8.7 on

page 36 for more information about test mode and

using the TEST menu.

NOTE:

All load cell calibration

settings are lost when the RESETCONFIGURATION

command is run.

5.1.4

Parameter Setting Commands

Parameter setting commands allow you to display or

change the current value for a particular configuration

parameter (Tables 5-3 through Table 5-8).
Current configuration parameter settings can be

displayed in either setup mode or normal mode using

the following syntax:

command<ENTER>

Most parameter values can be changed in setup mode

only. Use the following command syntax when

changing parameter values:

command=value<ENTER>

where

value

is either a number or a parameter value.

Use no spaces before or after the equal (=) sign. If you

type an incorrect command, the display reads

??

.

Changes to the parameters do not take effect until you

exit setup mode.
For example, to set the motion band parameter to 5,

type the following:

MOTBAND=5D<ENTER>

Command

Function

DUMPALL

List all parameter values

VERSION

Write 390HE software version

RS

Reset the indicator

P

Write current displayed weight with units
identifier.
The P command returns the current
displayed weight value to the EDP port,
along with the units identifier. If the indicator
is in an underrange or overload condition,
the weight value is replaced with ######
(overload) or %%%%%% (underrange).

S

Write one frame of stream format

Table 5-2. EDP Reporting Commands

Command

Description

Values

GRADS

Graduations

1–999 999

ZTRKBND

Zero track band

OFF, 0.5D, 1D, 3D

ZRANGE

Zero range

1.9%, 100%

MOTBAND

Motion band

1D, 2D, 3D, 5D, 10D, 20D, OFF

OVRLOAD

Overload

FS+2%, FS+1D, FS+9D, FS

DIGFLTR1
DIGFLTR2
DIGFLTR3

Digital filtering

1, 2, 4, 8, 16, 32, 64

DFSENS

Digital filter cutout sensitivity

2OUT, 4OUT, 8OUT, 16OUT, 32OUT, 64OUT, 128OUT

DFTHRH

Digital filter cutout threshold

NONE, 2DD, 5DD, 10DD, 20DD, 50DD, 100DD, 200DD, 250DD

TAREFN

Tare function

BOTH, NOTARE, PBTARE, KEYED

Table 5-3. CONFIG EDP Commands