0 calibration, 1 front panel calibration – Rice Lake Survivor 390HE Hostile Environment Digital Indicator User Manual

Page 24

Advertising
background image

20

SURVIVOR 390HE Installation Manual

4.0

Calibration

The 390HE can be calibrated using the front panel, EDP commands, or the Revolution

configuration utility.

Each method consists of the following steps:

Zero calibration

Entering the test weight value

Span calibration

Optional rezero calibration for test weights using hooks or chains.

The following sections describe the calibration procedure for each of the calibration methods.

Figure 4-1. Calibration (CALIBR) Menu

4.1

Front Panel Calibration

To calibrate the indicator using the front panel, do the

following:

1. Place the indicator in setup mode (display

reads

CONFIG

) and remove all weight from the

scale platform. If your test weights require

hooks or chains, place the hooks or chains on

the scale for zero calibration.

2. Press until the display reads

CALIBR

(see

Figure 4-1). Press to go to zero calibration

(

WZERO

).

3. With

WZERO

displayed, press to calibrate

zero. The indicator displays

*CAL*

while

calibration is in progress. When complete, the

A/D count for the zero calibration is

displayed. Press again to save the zero

calibration value and go to the next prompt

(

WVAL

).

Figure 4-2. Editing Procedure for Numeric Values

4. With

WVAL

displayed, place test weights on

the scale and press to show the test weight

value. Use the procedure shown in Figure 4-2

to enter the actual test weight, then press to

save the value and go to span calibration

(

WSPAN

).

5. With

WSPAN

displayed, press to calibrate

span. The indicator displays

*CAL*

while

calibration is in progress. When complete, the

A/D count for the span calibration is

displayed. Press again to save the span

calibration value and go to the next prompt

(

REZERO

).

6. The rezero function is used to remove a

calibration offset when hooks or chains are

used to hang the test weights.

If no other apparatus was used to hang the

test weights during calibration, remove

the test weights and press to return to

the CALIBR menu.

If hooks or chains were used during

calibration, remove these and the test

weights from the scale. With all weight

removed, press to rezero the scale. This

function adjusts the zero and span

calibration values. The indicator displays

*CAL*

while the zero and span calibrations

are adjusted. When complete, the adjusted

A/D count for the zero calibration is

displayed. Press to enter the value, then

press to return to the CALIBR menu.

7. Press until the display reads

CONFIG

, then

press to exit setup mode.

XXXXXXX

XXXXXXX

WZERO

*CAL*

Display and edit

zero calibration

A/D count value

Display and edit

test weight value

WVAL

*CAL*

Display and edit

span calibration

A/D count value

WSPAN

VERS

PROGRM

PFORMT

SERIAL

CALIBR

CONFIG

FORMAT

*

CAL*

Press Enter to

remove offset from

zero and span

calibrations

REZERO

When editing numeric values, press

or

to change th e

digit selected. Press

or

to increment or decrement the

value of the selected digit. Press

to save the value.

2

0 0 0 0 0

After entering the value, press

or

to change the decimal

point location. Press

again to save the value and exit the

parameter.

2

0 0 0 0 0