SURVIVOR 390HE Installation Manual

2.0

Installation

This section describes procedures for connecting load

cell and serial communications cables to the 390HE

indicator. Assembly drawings and parts lists are

included for the service technician.

Use a wrist strap to ground yourself and protect
components from electrostatic discharge (ESD) when
working inside the indicator enclosure.

This unit uses double pole/neutral fusing which could
create an electric shock hazard. Procedures requiring
work inside the indicator must be performed by
qualified service personnel only.

The supply cord serves as the power disconnect for the
390HE. The power outlet supplying the indicator must
be installed near the unit and be easily accessible

2.1

Unpacking and Assembly

Immediately after unpacking, visually inspect the

390HE to ensure all components are included and

undamaged. The shipping carton should contain the

indicator, this manual, and a parts kit. If any parts

were damaged in shipment, notify Rice Lake

Weighing Systems and the shipper immediately.
The parts kit contains the items listed below:

Identification label.

Capacity and identification labels.

Identification label (PN 49958) includes

replacement overlay decals for indicators

using units other than pounds and kilograms.

3-position (PN 15888) and 7-position (PN

23165) pluggable terminal blocks for load cell

and serial communications connectors.

2.2

Enclosure Disassembly

The indicator enclosure must be opened to connect

load cell and communications cables. Power-off the

indicator and remove the screws that secure the

enclosure cover. Open the cover; the CPU/display

board is mounted to the enclosure cover.

2.3

Cable Connections

The 390HE provides three cord grips for cabling into

the indicator: one each for the load cell cable, serial

communications, and the AC power cord.

Figure 2-1. 390HE Cord Grip Assignments

2.3.1

Load Cells

To attach cable from a load cell or junction box,

remove connector J1 from the board. The connector

plugs into a header on the board. Connect cable from

the load cell or junction box through the load cell

cable cord grip to connector J1 as shown in Table 2-1.

If using 6-wire load cell cable (with sense wires),

remove jumpers JP1 and JP2 before reinstalling

connector J1 (see Figure 2-2). For 4-wire installation,

leave jumpers JP1 and JP2 on.
When connections are complete, reinstall connector J1

on the board.

Caution

J1 Pin

Function

1

+SIG

2

–SIG

3

+SENSE

4

–SENSE

5

SHIELD (nsee NOTE below)

6

+EXC

7

–EXC

NOTES:

• J1 SHIELD wire connection not used. Use ring terminal

to attach shield wire to grounded CPU board mounting
screw next to the J1 connector.

• For 6-wire connections, remove jumpers JP1 and JP2.

Table 2-1. J1 Pin Assignments

T

D I G I T A L W E I G H T I N D I C A T O R

MODE

ENTER

SAMPLE

PRINT

I/O

ZERO

GROSS

NET

TARE

UNITS

PRINT

T

B/G

Units

POWER

lb

kg

PC

AC Power

Serial Communications

Load Cell Cable