420HE Installation Manual - Calibration

27

4.0

Calibration

The

420HE

can be calibrated using the front panel, EDP commands, or the

Revolution

®

configuration utility. Each

method consists of the following steps:

Zero calibration

Entering the test weight value

Span calibration

Optional rezero calibration for test weights using hooks or chains.

The following sections describe the calibration procedure for each of the calibration methods.

XXXXXXX

XXXXXXX

XXXXXXX

XXXXXXX

WZERO

WV AL

WSPAN

ALGOUT

DIGI N

SETPNT

PROGRM PFORMT

SERIAL

CALIBR

CONFIG FORMAT

REZERO

VERS

Figure 4-1. Calibration (CALIBR) Menu

4.1

Front Panel Calibration

To calibrate the indicator using the front panel, do the
following:

1. Place the indicator in setup mode (display

reads

CONFIG

) and remove all weight from the

scale platform. If your test weights require
hooks or chains, place the hooks or chains on
the scale for zero calibration.

2. Press

until the display reads

CALIBR

(see

Figure 4-1). Press to go to zero calibration
(

WZERO

).

3. With

WZERO

displayed, press to calibrate

zero. The indicator displays

*CAL*

while

calibration is in progress. When complete, the
A/D count for the zero calibration is displayed.
DO NOT adjust this value after WSPAN has
been set!
Press again to save the zero
calibration value and go to the next prompt
(

WVAL

).

4. With

WVAL

displayed, press to show the

stored calibration weight. Use the procedure
shown in Figure 3-4 15 to enter the actual
value of the test weights to be used for the span
calibration. Press to store the entered

WVAL

value and go to span calibration (

WSPAN

).

5. With

WSPAN

displayed, place test weights on

the scale and press to calibrate span. The
indicator displays

*CAL*

while calibration is in

progress. When complete, the A/D count for
the span clibration is displayed. Press again
to store the span calibration value and go to the
next prompt (

REZERO

).

6. The rezero function is used to remove a

calibration offset when hooks or chains are
used to hang the test weights.

• If no other apparatus was used to hang the test

weights during calibration, remove the test
weights and press to return to the CALIBR
menu.

• If hooks or chains were used during

calibration, remove these and the test weights
from the scale. With all weight removed, press

to rezero the scale. This function adjusts the

zero and span calibration values. The indicator
d i s p l a y s

* C A L *

w h i l e t h e z e r o a n d sp a n

calibrations are adjusted. When complete, the
adjusted A/D count for the zero calibration is
displayed. Press to save the value, then press

to return to the CALIBR menu.

7. Press

until the display reads

CONFIG

, then

press to exit setup mode.

4.2

EDP Command Calibration

To calibrate the indicator using EDP commands, the
indicator EDP port must be connected to a terminal or
personal computer. See Section 2.3.3 on page 8 for
EDP port pin assignments; see Section 5.0 on page 29
for more information about using EDP commands.

Once the indicator is connected to the sending device,
do the following:

1. Place the indicator in setup mode (display

reads

CONFIG

) and remove all weight from the

scale platform. If your test weights require
hooks or chains, place the hooks or chains on
the scale for zero calibration.

2. Send the WZERO EDP command to calibrate

zero. The indicator displays

*CAL*

while

calibration is in progress.