3 revolution® calibration, 4 more about calibration, 1 adjusting final calibration – Rice Lake SURVIVOR 420HE Hostile Environment - Installation Manual User Manual

Page 32: 2 a/d counts, Adjusting final calibration, A/d counts

Advertising
background image

28

420HE Installation Manual

3. Place test weights on the scale and use the

WVAL command to enter the test weight value
in the following format:

WVAL=nnnnnn<CR>

4. Send the WSPAN EDP command to calibrate

span. The indicator displays

*CAL*

while

calibration is in progress.

5. To remove an offset value, clear all weight

from the scale, including hooks or chains used
to hang test weights, then send the REZERO
EDP command. The indicator displays

*CAL*

while the zero and span calibrations are
adjusted.

6. Send the KUPARROW EDP command to exit

setup mode.

4.3

Revolution® Calibration

To calibrate the indicator using

Revolution

, the

indicator EDP port must be connected to a PC running
the

Revolution

configuration utility.

Use the following procedure to calibrate the indicator:

1. Select

Calibration Wizard

from the

Revolution

tools menu.

2.

Revolution

uploads calibration data from the

indicator then presents the information in a
display like that shown in Figure 4-2.

Figure 4-2.

Revolution Calibration Display

3. Enter the

Value of Test Weight

to be used for span

calibration then click

OK

.

4. The Zero Calibration dialog box prompts you

to remove all weight from the scale. Clear the
scale and click

OK

to begin zero calibration.

Note

If your test weights require hooks or
chains, place the hooks or chains on the scale
for zero calibration.

5. When zero calibration is complete, the Span

Calibration dialog box prompts you to place
test weights on the scale for span calibration.
Place tests weights on the scale then click

OK

.

6. When calibration is complete, the

New Settings

fields of the Indicator Calibration display are
filled in. Click

Exit

to save the new values and

return to the

Revolution

main menu; to restore

the previous calibration values, click

Restore

Settings

.

4.4

More About Calibration

The following topics provide additional information
about compensating for environmental factors
(Section 4.4.1) and diagnostic information for
determining expected zero and span coefficients.

4.4.1

Adjusting Final Calibration

Calibration may be affected by environmental factors
including wind, vibration, and angular loading. For
example, if the scale is calibrated with 1000 lb, a strain
test may determine that at 2000 lb the calibration is 3 lb
high. In this case, final calibration can be adjusted by
tweaking WVAL to 998.5 lb. This adjustment provides
a linear correction of 1.5 lb per 1000 lb.

To adjust the final calibration, return to the

WVAL

prompt and press to show the test weight value.
Press or to adjust calibration up or down. Press
to save the value, then press to return to the CALIBR
menu.

4.4.2

A/D Counts

Table 4-1 lists the ideal A/D counts that result from
input signals of 0–3.0 mV with zero deadload. Actual
values will typically be higher than the values shown in
Table 4-1 but the ideal values can be used when
calibrating the indicator with no attached scale.

Table 4-1. Ideal A/D Raw Counts

Input Signal (mV)

Raw A/D Counts

0.0 mV/V

167,840

0.5 mV/V

335,613

1.0 mV/V

503,377

1.5 mV/V

671,143

2.0 mV/V

838,908

2.5 mV/V

1,100,668

3.0 mV/V

1,174,446

Note

When Raw A/D Count is displayed, the six
most significant digits appear. Scroll left or
right to see the full value.

Advertising