Limit function, 1 limit function setup – Rice Lake TC Series, Rice Lake Weighing Systems - Operation Manual User Manual

Page 27

Advertising
background image

22 RL-TC Series Operation Manual

8. Limit Function

The limit function judges measurements according to a limit value saved in the balance.

The function shows the judgement result by displaying the [

] mark as either HI (excessive), OK (appropriate), or LO

(insufficient). This function is very useful when discriminating between conforming and nonconforming articles. It is
also useful when measuring a given constant quantity consecutively, in conjunction with a range of reference weights
defined by upper- and lower-limit values.

This function can be used in weight mode, counting mode, or percentage mode.

Limit value input methods

Either of the following two methods can be used in the different modes:

(1) Actual quantity setup method......... An actual sample is loaded on the balance and its weight saved as the limit

value.

(2) Numeric value setup method ......... The limit value is entered with a key stroke.

※The limit values entered are held in memory, even when the balance is powered down.

※The respective limit values for weight mode, counting mode, and percentage mode are set up independently.

Indication of judgment result

The [

] mark lights up as either HI, OK, or LO on the left side of the display, indicating the result of judgement.

Judgement Results

Upper/lower-limit setting

One-point setting

HI (excessive)

Upper-limit value < measurement
value

No indication

OK (appropriate)

Upper-limit value

≥ measurement

value

≥ lower-limit value

Limit value

≤ measurement value

LO (insufficient)

Lower-limit value > Measurement
value

Limit value > Measurement value

8.1 Limit Function Setup

1

Invoking the function

Press and hold down the Function key.

Release the key when [

Func

] is

displayed.
The display changes to the function
setup and the first item is displayed.

Advertising