24 RL-TC Series Operation Manual

8.2 Setup of Limit Values by Actual Quantity Loads

1

Starting the limit function

Press and hold down the Target key.

Release the key when [

L

.

SEt

] is

displayed.
The currently set lower-limit value
flashes.

2

Loading the sample for the lower-limit value

Load the sample of the lower-limit value
on the measurement pan.

3

Saving the lower-limit value

Press the Memory key.

After the lower-limit value has been
saved, the balance displays it briefly and
proceeds to the following setup.
※If One-point setup was chosen, the

setup is complete.

The display now changes to
[

H. SET

], indicating that the upper-limit

value can be set.
The currently set upper-limit value
flashes.

4

The upper-limit

value setup

5

Loading the sample of the upper-limit value

Load the sample of the upper-limit value
on the measurement pan.

6

Saving the upper-limit value

Press the Memory key.

After saving the upper-limit value, the
balance displays it briefly and terminates
the setup.