Torbal AGS100 250C PRO Manual User Manual

Page 13

Advertising
background image

13

1. If the Analyzer is in weighing mode, press the MODE RETURN key and enter the Moisture Analysis

Mode.

2. Press the SET (2) key to enter the Moisture Analyzer main menu. Configuration parameters will be

displayed: Mode, Drying Temp, Samples Quantity, Sampling Interval, Drying Time, and Save
Parameters.

8.1. Mode

1. Use the “<>” navigation key to select Mode and press the ENTER key.

2. The current mode will be displayed. Use the “<>” navigation key to alternate between the Time Mode

and the Short Mode. Once a desired mode is displayed, press the ENTER key to confirm the selection.

Short Mode – the analysis will be terminated automatically when the loss in moisture between the
samples taken during the analysis is less than 2mg. This mode requires that the Sample Quantity
(Section 8.4) and the Sampling Interval (Section 8.5) are defined.

Time Mode – the analysis is terminated when the defined Drying Time (Section 8.6) is reached.


Chapter 8: Configuration and Parameter Setting for Moisture Analysis

Advertising
This manual is related to the following products: