Chapter 9: rap configuration – Torbal AGS100 250C PRO Manual User Manual

Page 18

Advertising
background image

18

To enter Statistical Configuration, press the F2 key (STAT). The following STAT options will be
displayed:

Drying chart – the Drying Chart is especially useful during initial testing of new materials; however, it

should be turned OFF during routine analysis. It is very helpful in defining parameters such as drying
temperature, sampling interval, and drying time, as the shape of the drying cycle is clearly displayed.

Product name – this setting allows the User to assign a name of the product or sample being analyzed.

The name will be printed on the Drying Report printout (maximum 20 characters total).

Executive – this setting allows the User to assign a name of the person performing the analysis. The

name will appear on the Drying Report printout (maximum 20 characters total).

Notes – this setting allows the User to enter important notes that may be required for the analysis. The

notes will appear on the Drying Report printout (maximum 240 characters total).

Transmission – when turned ON, this function sends the sample weight to the printer port at the end of

each sampling interval. This function can be very useful during initial sample evaluations or for sending
data to an external computer for more detailed analysis.


To set and configure the Functions, follow the steps below:

1. Use the Navigation keys to select a STAT function you would like to configure.

2. Once the function is selected, press the ENTER key.