Torbal AGS100 250C PRO Manual User Manual

Page 37

Advertising
background image

37

9. When the calibration weight is placed on the pan, the scale will automatically begin the calibration

process.

10. When calibration is complete, the unit requests that the External Load be removed. When the Load is

removed from the unit, the screen requests a waiting period, which is followed by a screen showing
dashes (- - - - - -). After completing the calibration process, the unit will return to the Calibration Menu.

11. To get a Report Printout of the calibration results, use the navigation keys to select 3. REPORT

PRINTOUT and press the ENTER key. Calibration Data will be sent to the external printer.


12. To exit the Calibration menu, select EXIT and press the ENTER key.