When the Analyzer is initially installed, it must be calibrated to ensure accurate weighing results.
Calibration should be performed periodically or whenever the unit is moved to a different location. Before
calibrating the Analyzer, have the appropriate calibration weight available.

Note: In order to perform Calibration, the Analyzer must be in the Weighing Mode.

1. Press the Menu key to enter the Main Menu of the weighing mode.


2. Use the navigation keys to select CALIBRATION and press the ENTER key

3. To change the calibration mass, use the navigation keys to select EXTERNAL LOAD and press ENTER.



4. Use the navigation keys to select the allowed external loads for calibration. The load equal to full scale

range is recommended. Be sure the weight used for the External Load is of sufficient accuracy for your
purpose. Once the Calibration mass has been selected press the ENTER key to accept the External Load
selection.