Brookfield YR-1 Rheometer User Manual

Page 10

Advertising
background image

Brookfield Engineering Labs., Inc.

Page 10

Manual No. M02-215-B0412

NOTICE TO CUSTOMERS:

This symbol indicates that this product is to be recycled at an appropriate collection center.

Users within the European Union:

Please contact your dealer or the local authorities in charge of waste management on how to

dispose of this product properly. All Brookfield offices and our network of representatives and

dealers can be found on our website: www.brookfieldengineering.com

Users outside of the European Union:

Please dispose of this product according to your local laws.

I.4 Installation

1. Assemble the Model S Laboratory Stand (refer to assembly instructions in Appendix F).

2. Connect the RTD probe to the socket on the rear panel of the YR-1.

3. The Rheometer must be leveled. The level is adjusted using the two leveling screws on the

base. Adjust so that the bubble level on top of the YR-1 is centered.

Note: Check level periodically during use.

4. Remove the white shipping cap, which secures the spindle coupling nut on the Rheometer

to the pivot cup.

5. Make sure that the AC power switch at the rear of the YR-1 is in the OFF position. Con-

nect the power cord to the socket on the back panel of the instrument and plug it into the

appropriate AC line.

The AC input voltage and frequency must be within the appropriate range as shown on the

nameplate of the viscometer.

Note: The YR-1 must be earth grounded to ensure against electronic failure!!

6. Turn the power switch to the ON position and allow the YR-1 to warm up for 10 minutes

before performing the autozero.

7. If appropriate, attach the interconnecting cable to the serial port to connect the YR-1 to a

PC or printer.

8. If appropriate, attach the interconnecting cable to the parallel port to connect the YR-1 to a

printer.

9. If appropriate, attach the interconnecting cable to the analog (serial) port to connect the

YR-1 to a chart recorder.

Advertising