Brookfield YR-1 Rheometer User Manual

Page 30

Advertising
background image

Brookfield Engineering Labs., Inc.

Page 30

Manual No. M02-215-B0412

#5 Sample Tst

↑↓

START TEST

Figure III-7

Note:

While the top line is the active selection line, the program number and name will

always be flashing.

Pressing the · or ‚ key scrolls through the program numbers (0-9) and displays the corresponding

program name. While scrolling, if the slot scrolled to has no program in it, EMPTY is displayed

next to the program number, and pressing the Enter key has no effect.

Pressing the ESC key with this screen displayed causes the screen in Figure III-5 to again be

displayed (i.e. go back one level).

Once the sample is prepared and the Enter key is pressed with a valid program displayed, the

following screen is displayed:

USE SPINDLE #71

PRESS ENTER

Figure III-8

where the two (2) digit spindle number is displayed on the top line. Attach the appropriate

spindle and lower it into the sample before continuing.

Pressing the Enter key with Figure III-8 displayed performs the following:

• The test is started.
• The pre-test information (test parameters) is sent to both the serial (RS-232) port and to an

attached printer via the parallel printer port. See section II.3 for a description of the pre-test

information sent to the printer.

• The program number being run is recorded so that it will be the first program shown the

next time that START TEST is selected from the Default Screen (Figure III-5).

Pressing the ESC key at this screen causes Figure III-7 to be displayed.

After ensuring the appropriate spindle is attached and pressing the Enter key, the various Run

screens are displayed. The top line of the Run screens always displays the program and spindle

number as follows:

TEST #5 SPDL 71

---------------------------

Figure III-9