Brookfield YR-1 Rheometer User Manual

Page 15

Advertising
background image

Brookfield Engineering Labs., Inc.

Page 15

Manual No. M02-215-B0412

these fields until the check mark appears enables these parameters. Clicking the check box to

the left of these fields until the check mark disappears disables these parameters.

Clicking the New button (or selecting New from the File menu) sets all parameters to their

default values.

All open, save, and print operations selected while this page is displayed will open, save, and

print test parameters, but NOT data.

II.3.1

Program Number

This is the number of the memory slot in the YR-1 Rheometer to which the test parameters will

be loaded. There are ten (10) slots, numbered from 0 through 9.

II.3.2

Program Name

This descriptive user-supplied name is loaded into the memory slot in the YR-1 Rheometer with

the test parameters. A maximum of ten (10) alphanumeric characters may be used for this name.

II.3.3 Spindle

Spindle Number

A two (2) digit code representing the spindle number used for the test must be selected. See

Appendix A for more information regarding spindles for use with the YR-1 Rheometer. Selec-

tion of the appropriate spindle code is important to ensure correct stress calculations.

Primary

Immersion

Mark

Secondary

Immersion

Mark

Figure II-2

Immersion Mark

Each vane spindle has two (2) immersion marks. The primary immersion mark is located on the

spindle shaft. Normally, the spindle should be inserted so that the sample reaches this mark. If

the sample container is too small to allow the spindle to be inserted to this mark, the secondary

Advertising